AMP-8


AMP-8  [for use with HFM-6 and HFM-7 series - factory adjusted (specified by customer) gain setting]
specification sheet also available in PDF file
 
 

Vatell AMP-11
 

Advantages:

- Customer-specified gain to fit individual applications
- Indefinite Continuous Use Possible with 5V AC-to-DC adapter that plugs into the wall
- Compact Size
- Simple Configuration Allows Easy, Straight-forward Use
- Input cables 5V AC-to-DC Adapter Included

Specification:
.
Gain is adjustable.  Gains in table were chosen for convenince
Heat Flux Channels
Temperature (RTS) Channel 
Gain Settings
Customer-specified
Customer-specified
Gain Accuracy (%)    
Gain = 1
± 0.6
± 0.6
Gain = 100
± 1.5
± 1.5
Gain = 200, 500
± 1.5
± 1.5
Gain = 1000
 ± 2.1
 
Gain = 5000
± 3.6
 
Bandwidth
 (At some selected gains)
 (At some selected gains)
Gain = 1
 1 MHz
 1 MHz
Gain = 100
 150 kHz
150 kHz 
Gain = 200
 
100 kHz
Gain = 500
 50 kHz
50 kHz 
Gain = 1000
 25 kHz
 
Gain = 5000
 5 kHz
 
Input Impedance
 109 Ohms
 109 Ohms
Input Noise
 0.2 µV
 0.2 µV
Full Scale Output
 6 Volts
6 Volts 

Principle of Operation:
The AMP-8 is a 2 channel signal-conditioning instrumentation amplifier designed to be used with a Vatell Corporation Heat Flux Microsensor (HFM).  The HFM consists of two sensors on the same surface.  The Heat Flux Sensor (HFS) measures heat flux flowing through the sensor surface.  The Resistance Temperature Sensor (RTS) measures the temperature at the face of the sensor.  One channel of AMP-8 amplifies signals from the HFS portion of the HFM.  The other channel excites the RTS portion of the HFM with a 100 microamp current source and then amplifies the resulting voltage signal.

Construction:
The AMP-8 has a four-pin female Lemo connector to interface with the Heat Flux Microsensor.  Male BNC connectors are provided for amplified heat flux and temperature output signals.  The unit is powered with a 5 V AC-to-DC adapter that plugs into the wall.  Gains for heat flux and temperature channels are specified by the customer.  The channels are then set and zeroed   by the customerl.

Dimensions:

1.25 x 1.5 x 4 inches
Weight: 6oz
 
 