Absorbed vs. Incident Heat Flux Calibrations
Radiant heat flux that is incident upon a heat flux transducer is always partially reflected. The signal produced by the transducer represents only the absorbed heat flux, since reflected energy does not enter the signal generation process. If the absorptivity of the transducer surface were exactly known, calibrations could be in terms of absorbed heat flux. However, the absorptivity is not accurately known. Furthermore, calibrations must be in terms of something which can be accurately measured. This is the incident heat flux.
All known standards for heat flux; those based on radiometric measurements
and those based on temperature measurements, are in terms of incident heat
flux. To calibrate a heat flux transducer, we illuminate it with
a source of radiant energy whose incident heat flux at the surface of the
transducer is accurately known. We measure the output signal of the
surface of the transducer with one or more values of incident heat flux.
The sensitivity and scale factor of the transducer are then calculated,
based on incident heat flux and the signal it produces.
As long as the coating on the surface of the transducer is not changed, the relationship between incident heat flux and output signal will remain the same. However, if the coating is removed, worn away, sooted up, or if the transducer is re-coated, the calibration will no longer be valid. This is because the absorptivity of the coating, while not known accurately, is a factor in the sensitivity of the transducer.
When a heat flux transducer is used for convective or conductive measurements,
there is no reflection process. All of the heat flux that passes
through the front surface of the transducer also passes through the measuring
element. It would be erroneous to use the sensitivity and scale factor
values derived from a radiative calibration for these two modes of heat
transfer. Instead, the values of sensitivity and scale factor must
be modified. The radiative sensitivity is divided by the emissivity
(» absorptivity) to derive a conductive/convective sensitivity.
The scale factor of the transducer is multiplied by the emissivity.
The sensitivity and scale factor derived in this way are correct even if
the surface of the transducer is changed, because they are based
on the original calibration
and the original coating.